7 | Warning and error messages INFICON
98 / 152 UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109)
Type Notification Possible sources of
error
Remedy
W125 I/O module no longer
connected
Connection to I/O
module interrupted
Check the connection to the I/O module
Replace the connection cable to the I/O
module
I/O module defective Replace the I/O module
I/O module connection
on the device defective
Contact customer service
W127 Wrong bootloader version The bootloader is not
compatible with
application
Contact customer service
E129 EEPROM contains data
from wrong device class
The software of the
basic unit does not
match the EEPROM
Contact customer service
The EEPROM does not
match this device class
Contact customer service
W130 Sniffer line not connected The sniffer line cannot
be addressed by the
device
Check the connection between sniffer line
and the device (disconnect and
reconnect)
Contact customer service
Sniffer line defective Use another sniffer line, if possible
Contact customer service
W132 Sniffer line type (XL) not
allowed
A sniffer line type (XL)
cannot be used with
this device
Replace the sniffer line with a sniffer line
type supported by the device
W151 No communication with
operating unit
A software update or a
parameter reset has
been executed
Confirm the warning message
Check that the message does not appear
when you restart the device
Internal connection
problem between the
basic unit and the
operating unit
Contact customer service
W152 No communication with
control panel
Internal connection
problem between the
basic unit and the
control panel
Contact customer service