INFICON Warning and error messages | 7
UL3000-Fab-ULTRA-Operating-instructions-iinb73en1-09-(2109) 99 / 152
Type Notification Possible sources of
error
Remedy
W153 Operating unit software
version is obsolete
A more up-to-date
operating unit software
is available. For
trouble-free operation,
it is recommended to
update the operating
unit software.
Contact the customer service for the
latest operating unit software
W171 CU1000 not supported A CU1000 cannot be
used with this device
Disconnect the CU1000 from this device
W201 Voltage U24V_MSB too low Malfunction of 24V
power supply unit
Contact customer service
Short circuit or
overload in the 24V
supply
Contact customer service
W202 Voltage U24V_MSB too high Malfunction of 24V
power supply unit
Contact customer service
W203 24V valve supply voltage out
of range
Malfunction of a valve Contact customer service
Short circuit or
overload in the 24V
valve supply
Contact customer service
W206 24V operating unit supply
voltage out of range
Malfunction of
operating unit
Contact customer service
Short circuit or
overload in the 24V
operating unit supply
Contact customer service
W207 12V internal supply voltage
out of range
Short circuit or
overload in the 12V
supply
Contact customer service
W208 24V fan supply voltage out
of range
Malfunction of a fan Contact customer service
Short circuit or
overload in the 24V fan
supply
Contact customer service
W209 24V DIV1 supply voltage out
of range
Short circuit or
overload in the 24V
DIV1 supply
Contact customer service
W210 24V DIV2 supply voltage out
of range
Short circuit or
overload in the 24V
DIV2 supply
Contact customer service