Remote Control 83
8.3 Load setting commands
Function Command Operating/setting range Memo
Load setting CC{NR2} Constant curr.A When setting load consditions,
CR{NR2} Constant registorΩ〕 mode and setting values are setup
CV{NR2} Constant volt.V at the same command.
CP{NR2} Constant powerW
CX{NR2} Curr. value[A] when ext. ctrl volt is 10V.
Short
Curr. range becomes High range
CD{NR1} range: 02
When CD command is issued,
0stop/shift dynamic operation dynamic seettings become activate.
1start dynamic operation (loop)
2start dynamic operation (single)
Dynamic DP{NR1} range: 0n(max:16) Section where DC and DTcommands are
1define "TIME A" and section "CC-1". are efficitve is defined.
2define "TIME B" and section "CC-2"
ndefine "TIME n" and section "CC-n" (*2)
DC{NR2} curr. value(A) in section. efficitve for the section defined by
DT{NR2} time period (msec) of section. DP command
DR{NR2} set load resp. (slew rate) (A/μsec) settable regardless of DP setting
Curr. setting range RC{NR1} range : 02 *1 load current range
0auto range
1High range
2Low raneg
Volt. setting range RV{NR1} range: 12 *1 load volt. range
1High range This setting influences CR and CV
2Low raneg settings.
Output control SW{NR1} range: 01 Load OFF becomes no load condition
0load OFF inactivating load settings.
1load ON
Load resp. (slew rate) GC{NR2} resp. time [μ/A] in CC setting Valid for CC and CV setting only.
GV{NR2} resp. time [%] in CV setting or max default value in other modes.
Volt. limit setting LV{NR2} limit voltage active except 0V setting
Curr. limit setting LS{NR2} Curr value[A] up to max. rated value. No limit for load curr. setting range.
Booster setting BS{NR1}
Range: 0-2
0:Normal
1: Master
2: Multi
To set Master/Multi, present setting has
to be in Normal.
range : 01
0 : OFF
1 : ON
Notes:
  { } : you can NOT mot in this parenthesis.
    NR1・・・・・・ integral value
    NR2・・・・・・ real value
*1: Setting range and the meas. range of load section can not be setup independently but should be set as identical..
*2: Pushing load on button, a single sequence mode is run.
CS
A setup of "CD1 -> CD2" or "CD2 ->
CD1" cannot be performed.
BIAS power control
BI{NR1}
Cannot turn BIAS power on while load is
on.