58 Operations for Load Mode
5.12 Dynamic Load mode (DYNAMIC)
This mode repeats multiple loading conditions.
All loading conditions of CC, CV, CP, EXT and SHORT modes
can be set. However each loading condition has to be in a same
loading mode.
Max.16 patterns can be set.
(1) Multiple load condition setting
The DYNAMIC loading can be achieved by multiple loading conditions.
CC1 and CC2 have the following parameters which can be setup independently each other.
Fig. 5-12-1 DYNAMIC loading
・Load value
You can setup these values same as in each mode selected
・Effective time
from 1μs to 60s (1μs resolution)
・Response speed (slew rate)
Above parameters can be setup independently from each other.
・ The accuracy of the Effective time becomes coarser as the effective time becomes longer,
even though the resolution is not changed.
・ In the dynamic mode operation, trigger signals are outputted from the TRIG OUT output on
the front panel in the timing of circled edges in the Fig. 5-11-1.
For more information, refer to →”2.5 TRIG OUT “.