Load Edge Series
Alarms 75
Chapter 7 Alarms
Alarms of this product are explained.
In this product, when an alarm occurs, Load is automatically set to OFF, with beep sound rung
and error message displayed.
7.1 Over current limit
To protect load section, load current of this product can be limited. When you setup the
current Limit current will be limited with the 110% (typ.) of setting value. The default
value is 67.5A.
Table. 7-1-1 Default current limit setting
For the operation of setting current limit, refer to → “ 5.4 Current limit setting ”.
Fig. 7-1-1 Display of over current alarm
・ When the limit value is set less than 15 % (Typical) of range (about 10A in High
range 67.5A, about 1.1A in Low range 7.5A) the accuracy of the limit current will
be deteriorated.
・ Transient peak current over current limit may be drawn when in CV mode.
・ Never set current value more than 100 times than its limit value.