18-16 SCPI Command Reference 2400 Series SourceMeter
®
User’s Manual
[:SOURce[1]]
:SWEep Configure SWEep source mode:
:SPACing <name> Select sweep spacing type (LINear or
LOGarithmic).
LINear
:SPACing? Query sweep spacing.
:POINts <n> Specify number of sweep points (2 to
2500).
2500
:POINts? Query number of points in sweep.
:DIRection <name> Sweep from start to stop (UP) or from stop
to start (DOWN).
UP
:DIRection? Query sweep direction.
:RANGing <name> Select source ranging mode (BEST,
AUTO, or FIXed).
BEST
:RANGing? Query source ranging mode.
:LIST Configure LIST source mode:
:CURRent <NRf> Create list of I-Source values.
1
:APPend <NRf> Add I-Source values to end of list.
:POINts? Query number of source values in list.
:STARt <n> Specify list sweep start point. 1
:STARt? Query list sweep start point.
:CURRent? Query I-Source list.
:VOLTage <NRf> Create list of V-Source values. 1
:APPend <NRf> Add V-Source values to end of list.
:POINts? Query number of source values in list.
:STARt <n> Specify list sweep start point. 1
:STARt? Query list sweep start point.
:VOLTage? Query V-Source list.
:MEMory Configure Source Memory Sweep:
:SAVE <n> Save settings at memory location (1 to 100).
:RECall <n> Recall settings from memory (1 to 100). 1
:POINts <n> Specify number of sweep points (1 to 100).
:POINts? Query number of sweep points.
:STARt <NRf> Specify start location for Source Memory
Sweep (1 to 100).
1
:STARt? Query start location.
:PULSe Model 2430 — Configure pulse for Pulse
Mode:
:WIDTh <n> Set pulse width (0.00015 to 0.005 sec). ✓✓
:WIDTh? Query pulse width.
:DELay <n> Set pulse delay (0 to 9999.999 sec). ✓✓
:DELay? Query pulse delay.
1. The parameter values for this command are provided with the detailed command reference information. Detailed information on
SCPI commands follows the last command summary table in this section.
Table 18-7 (continued)
SOURce command summary
Command Description
Default
parameter SCPI
Source
memory
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Keithley SourceMeter 2420

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2420 and is the answer not in the manual?

Keithley SourceMeter 2420 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2420
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals