EasyManuals Logo

Keysight Technologies InfiniiVision MSO-X 3014A User Manual

Keysight Technologies InfiniiVision MSO-X 3014A
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
Math Waveforms 4
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide 81
• y = channel 1, 2, 3, or 4, or g(t) (internal arithmetic operation).
• co = arbitrary constant.
• i = data point index.
The integrate operator provides an Offset softkey that lets you enter a DC offset
correction factor for the input signal. Small DC offset in the integrate function
input (or even small oscilloscope calibration errors) can cause the integrate
function output to "ramp" up or down. This DC offset correction lets you level the
integrate waveform.