EasyManua.ls Logo

Keysight E4990A - Page 536

Keysight E4990A
669 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E4990A Help
526
:SENS:CORR:CKIT:OPEN:C
:SENS:CORR:CKIT:OPEN:G
:SENS:CORR:CKIT:OPEN:TABL
Equivalent Softkey
Cal > User Cal > Define Value > Open Model > Equivalent Circuit | f-Z Table
Cal > Fixture Compen > Define Value > Open Model > Equivalent Circuit | f-Z
Table
COM Command Syntax (Internal Control Only)
SCPI.SENSE(Ch).CORRECTION(Type).CKIT.OPEN.MODEL
:SENS<Ch>:CORR<Type>:CKIT:OPEN:TABL
Type
Command/Query
Syntax
:SENSe<Ch>:CORRection<Type>:CKIT:OPEN:TABLe <Value 1>, ... ,
<Value (n×3+1)>
:SENSe<Ch>:CORRection<Type>:CKIT:OPEN:TABLe?
Description
This command defines the table of open model for the user calibration and
fixture compensation.
(<type>: User calibration=1, Fixture Compensation=2)
Variable
Parameter Value
Description
Open standard definition table:
Indicates the array data of 1 + NOP (number of
standard data points) × 3. Where n is an integer
between 1 and NOP.
<value 1> :The number of standard data points you
want to set. Specify an integer ranging 1 to 1601.
<Value (n×3-1)> :Frequency at the n-th of standard
data.
<Value (n×3)> :Real part of data (complex number) at
the n-th standard data point.
<Value (n×3+1)> :Imaginary part of data (complex
number) at the n-th standard data point.
Data Type Variant type Array (Range)