Test Equipment and Setups
Test Equipment
Measurement Setups
Test Equipment
The test equipment recommended for the performance verification and adjustment procedures is
listed below. If the exact instrument is not available, substitute calibration standards of equivalent
accuracy.
Instrument Requirements Recommended Model Use
1
Digital Multimeter Resolution: 10 nV @ 1V
Readout: 8 1/2 digits
Accuracy: 20 ppm
Keysight 3458A V, C
Film capacitor 0.1 μF, 10%, 400 Vdc (for voltage ripple and
noise)
B32520E6104K000 or equivalent V
Current Shunt 15 A (0.1 Ω), 0.01%, TC=4ppm/ΩC Guildline 9230A/15R V, C
DC power supply 160 V, 10 A (for current sink verification) Keysight MP4361A or equivalent V
Load resistors 10 Ω, 1 kW, 1 % (for voltage ripple and noise) 4 x NH25010R00FE01, 250 W
or equivalent 10 Ω 1kW
V
Electronic load 160 V, 10 A, 1 kW minimum Keysight MP4361A module or equivalent V
GPIB controller Full GPIB capabilities Keysight 82350B or equivalent V, C
Oscilloscope Sensitivity: 1 mV
Bandwidth: 20 MHz
Keysight DSO6054A or equivalent V
RMS Voltmeter Bandwidth: 10 MHz Rhode and Schwartz URE3, Keysight
3458A, or equivalent
V
Differential Amplifier Bandwidth: 20 MHz LeCroy DA1855A, DA1850A, or equivalent V
Terminations 50 Ω BNC termination V
1
V=Verification; C=Calibration
Measurement Setups
Voltmeter
To ensure that the values read by the voltmeter during both the verification procedure and the
calibration procedure are not affected by the instantaneous measurement of the ac peaks of the
output current ripple, make several dc measurements and average them.
If you are using a Keysight 3458A DMM, you can set up the voltmeter to do this automatically. From
the instrument’s front panel, program 100 power line cycles per measurement. Press NPLC 100
ENTER. Additionally, turn on auto-calibration (ACAL) and the autorange function (ARANGE).
6 Verification and Calibration
168 Keysight MP4300 Series Operating and Service Guide