Reset State (*RST)
The power-on/reset state may differ from that shown below if you have enabled
power-on state recall mode from the States menu (see Instrument State Storage)..
Reset Settings
The following table shows the reset state. These parameters are reset to the indicated values at power-
on or after *RST.
SCPI Command *RST Settings
CALibrate:STATe OFF
CURRent 0
CURRent:LIMit 1.02% of rating
CURRent:LIMit:NEGative -1.02% of rating (MP4361A)
-0.51A (MP4362A)
CURRent:MODE FIXed
CURRent:PROTection 12.5 A (MP4361A)
10 A (MP4362A)
CURRent:PROTection:STATe OFF
CURRent:SASimulator:BWIDth:RANGe DEFAULT
CURRent:SASimulator:IMP 0.8% of nominal rating
CURRent:SASimulator:ISC 1% of nominal rating
CURRent:SAS:SCALe 100%
DIGital:OUTPut:DATA 0
FUNCtion VOLTage
LXI:IDENtify OFF
LXI:MDNS ON
OUTPut OFF
VOLTage 0
VOLTage:LIMit 102% of rating
VOLTage:LIMit:LOW 0
VOLTage:PROTection 120% of rating
VOLTage:SASimulator:VMP 0.8% of nominal rating
Keysight MP4300 Series Operating and Service Guide 153
5 SCPI Programming Reference

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight MP4300 Series and is the answer not in the manual?

Keysight MP4300 Series Specifications

General IconGeneral
BrandKeysight
ModelMP4300 Series
CategoryPower Supply
LanguageEnglish

Related product manuals