Test Record Forms - MPS family
Keysight MP4361A
MP4361A
Test Record
Report Number ______ Date__________
Test Description Min. Specs Results Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
-0.025
Vout - 0.025
159.895
Vout - 0.105
_______
_______
_______
_______
0.025
Vout + 0.025
160.105
Vout + 0.105
CV Ripple and Noise
peak-to-peak:
rms:
N/A
N/A
_______
_______
0.5
0.05
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
- 0.007
Iout - 0.007
9.983
Iout - 0.017
_______
_______
_______
_______
0.007
Iout + 0.007
10.017
Iout + 0.017
Current Sink Test
Negative current rating: - 10.017 _______ - 9.983
MP4361A UUT Settings Programmed Source Settings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Ripple and Noise:
Current Programming & Readback Min:
Current Programming & Readback High:
Current Rating (Isink):
Voltage priority: 0 V, 1 A
Voltage priority: 160 V, 1 A
Voltage priority: 100 V, 10.2 A
Current priority: 100 V, 0 A
Current priority: 100 V, 10 A
Current priority: 102 V, -10 A
not used
not used
not used
not used
not used
Voltage priority: 100 V, 10.2 A
6 Verification and Calibration
180 Keysight MP4300 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight MP4300 Series and is the answer not in the manual?

Keysight MP4300 Series Specifications

General IconGeneral
BrandKeysight
ModelMP4300 Series
CategoryPower Supply
LanguageEnglish

Related product manuals