FUNCtion Command
The Function command to set voltage or current priority only applies in Fixed mode
operation. It is not available in SAS and Table modes. Refer to the
SOUR:CURR:MODE command to specify the operating mode.
[SOURce:]FUNCtion CURRent|VOLTage, (@<chanlist>)
[SOURce:]FUNCtion? (@<chanlist>)
Sets the output regulation - voltage priority or current priority. In voltage priority mode, the output is
controlled by a constant voltage feedback loop, which maintains the output voltage at its
programmed setting. In current priority mode, the output is controlled by a constant current feedback
loop, which maintains the output current at its positive or negative programmed setting.
Refer to Priority Mode Tutorial for more information.
Parameter Typical Return
CURRent|VOLTage, *RST VOLTage CURR or VOLT
<chanlist> one or more channels
Sets the output regulation to current priority on channel 2: FUNC CURR, (@2)