EasyManua.ls Logo

Keysight MP4300 Series - Function Command

Keysight MP4300 Series
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FUNCtion Command
The Function command to set voltage or current priority only applies in Fixed mode
operation. It is not available in SAS and Table modes. Refer to the
SOUR:CURR:MODE command to specify the operating mode.
[SOURce:]FUNCtion CURRent|VOLTage, (@<chanlist>)
[SOURce:]FUNCtion? (@<chanlist>)
Sets the output regulation - voltage priority or current priority. In voltage priority mode, the output is
controlled by a constant voltage feedback loop, which maintains the output voltage at its
programmed setting. In current priority mode, the output is controlled by a constant current feedback
loop, which maintains the output current at its positive or negative programmed setting.
Refer to Priority Mode Tutorial for more information.
Parameter Typical Return
CURRent|VOLTage, *RST VOLTage CURR or VOLT
<chanlist> one or more channels
Sets the output regulation to current priority on channel 2: FUNC CURR, (@2)
5 SCPI Programming Reference
116 Keysight MP4300 Series Operating and Service Guide

Table of Contents

Related product manuals