Command Quick Reference
Some [optional] commands have been included for clarity. All settings commands have a
corresponding query. See the Syntax Conventions for SCPI.
CALibrate
:COUNt? Returns the number of times the unit has been calibrated.
:CURRent
[:LEVel] <range> , (@<channel>) Calibrates the current; the value selects the range.
:DATA <value> Enters the calibration value read by the external meter.
:DATE <"date">, (@<channel>) Enters the calibration date in nonvolatile memory.
:LEVel P1|P2|P3 Selects the calibration step.
:PASSword <value> Sets a numeric password to prevent unauthorized calibration.
:SAVE Saves the calibration constants in non-volatile memory.
:STATe 0|OFF|1|ON Enables or disables calibration mode.
:VOLTage
[:LEVel] <range>, (@<channel>) Calibrates the voltage; the value selects the range.
FETCh
[:SCALar]
:CURRent
[:DC]? (@<chanlist>) Returns the averaged measurement.
:VOLTage
[:DC]? (@<chanlist>) Returns the averaged measurement.
HCOPy
:SDUMp
:DATA? [JPG|PNG] Returns an image of the front panel display.
:DATA
:FORMat JPG|PNG Specifies the format for front panel images returned.
Keysight MP4300 Series Operating and Service Guide 21
1 Quick Reference