EasyManuals Logo

Keysight MP4300 Series User Manual

Keysight MP4300 Series
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #171 background imageLoading...
Page #171 background image
LETHAL VOLTAGES - Power modules generate output voltages up to 160 VDC!
Ensure that all instrument connections, load wiring, load and sense connections are
insulated.
Always turn off ac power when connecting or disconnecting any equipment on the
sense or output terminals of the unit.
Equipment Damage It is recommended to set the over-voltage protection function
of the instrument slightly higher than its operating point during the verification
procedure. This will prevent damage to any external equipment (electronic load,
differential amplifier) that may occur if the output voltage is accidentally
programmed higher than the prescribed voltage setting.
Turn the unit off or send a Reset command after completing the verification
procedure to return all instrument settings to their default values.
Verification Setups
Keysight MP4300 Series Operating and Service Guide 171
6 Verification and Calibration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight MP4300 Series and is the answer not in the manual?

Keysight MP4300 Series Specifications

General IconGeneral
BrandKeysight
ModelMP4300 Series
CategoryPower Supply
LanguageEnglish

Related product manuals