11 SYSTem Subsystem
104 Keysight U2722A/U2723A Programmer’s Reference
SYSTem:CDEScription?
Syntax
SYSTem:CDEScription?
This query is used when the SMU device is slotted into modular instrument
chassis. It identifies which slot is the module plugged into.
Remarks
It is only applicable when the U2722A/U2723A is used in the modular
instrument chassis.
If the U2722A/U2723A is not plugged into the chassis, then the returned query
value would be “+7, +0”, where “+7” is an invalid slot number and “+0” is the
chassis number.
Please refer to the U2781A Modular Instrument Chassis User's Guide for the
details.
Returned Query Format
<NR1>, <NR1>
Examples
The following shows the query when the U2722A/U2723A is slotted into slot 4 of
chassis #1.
SYST:CDES?
Typical Response: +4, +1
Below is the query when the U2722A/U2723A is plugged into slot 2 of chassis #2.
SYST:CDES?
Typical Response: +2, +2