SOURce Subsystem 9
Keysight U2722A/U2723A Programmer’s Reference 77
[SOURce:]VOLTage
Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <voltage>,
(@<ch>)
Sets the actual voltage magnitude of the unswept output signal in terms of the
present operating units. The units are set to the default value, or alternately to a
different value under the UNIT subsystem. The AMPLitude may be used to specify
the level for either a time varying or non-time varying signal.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? (@<ch>)
Queries the sourced voltage.
Parameters
Returned Query Format
<NR3>
Examples
The following command sets the output voltage to 5 V on channel 1.
VOLT 5, (@1)
The following query returns the programmed voltage level on channel 1.
VOLT? (@1)
Typical Response: +5.000000E+00
See Also
[SOURce:]CURRent
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
Item Type Range of values Default value
<ch> NR1 1 through 3 Required parameter
<voltage> NRf The maximum value is dependent on the voltage range set Required parameter