Error Messages 14
Keysight U2722A/U2723A Programmer’s Reference 131
The following table shows the list of command errors.
Error Device-dependent errors (these errors set the Standard Event Status register bit #3)
+249 Over Temperature Occurred
Over temperature has been detected. Please remove the load or switch off the device.
+254 Memory List Full
The active memory lists is full. No commands can be inserted anymore.
Table 14-1 Device-dependent errors (continued)
Table 14-2 Command errors
Error Command errors (these errors set the Standard Event Status register bit #5)
–100 Command error
Generic syntax error.
–101 Invalid character
An invalid character was found in the command string. You may have inserted a character such as #, $, or % in
a command keyword or within a parameter.
–102 Syntax error
Invalid syntax was found in the command string. You may have inserted a blank space before or after a colon in
the command header, or before a comma.
–103 Invalid separator
An invalid separator was found in the command string. Check for proper usage of , ; :
–104 Data type error
The wrong parameter type was found in the command string. You may have specified a number where a string
was expected, or vice versa.
–105 GET not allowed
A group execute trigger is not allowed in a command string.
–108 Parameter not allowed
More parameters were received than expected for the command. You may have entered an extra parameter, or
you have added a parameter to a command that does not accept a parameter.
–109 Missing parameter
Fewer parameters were received than expected for the command. You may have omitted one or more
parameters that are required for this command.