10 STATus Subsystem
86 Keysight U2722A/U2723A Programmer’s Reference
STATus:OPERation[:EVENt]?
Syntax
STATus:OPERation[:EVENt]?
This query returns the value of the Operation Status Event register. The Event
register is a read-only register, which stores (latches) all events that are passed by
the Operation NTR and/or PTR filter. Reading the Operation Status Event register
clears it.
Remarks
The following table lists the bit definitions for the Operation Status registers.
Once a bit is set, it remains set until cleared by reading the event register or
the clear status (*CLS) command.
NOTE
For more information on the SCPI Status System, refer to Chapter 2, “Status
system diagram” on page 21.
Bit number Decimal value Definition
0 to 1 Not Used 0 0 is returned.
2 DTG1-tran 4 The transient system has received a trigger signal and operation is running in channel 1.
3 DTG2-tran 8 The transient system has received a trigger signal and operation is running in channel 2.
4 DTG3-tran 16 The transient system has received a trigger signal and operation is running in channel 3.
5 WTG1-tran 32 The transient system is waiting for a trigger in channel 1.
6 WTG2-tran 64 The transient system is waiting for a trigger in channel 2.
7 WTG3-tran 128 The transient system is waiting for a trigger in channel 3.
8 to 15 Not Used 0 0 is returned.