SOURce Subsystem 9
Keysight U2722A/U2723A Programmer’s Reference 55
[SOURce:]CURRent:RANGe
Syntax
[SOURce:]CURRent:RANGe <range>, (@<ch>)
This command sets the output current range. At *RST, low current range is
selected.
[SOURce:]CURRent:RANGe? (@<ch>)
Queries the selected current range.
Parameters
Returned Query Format
<CRD>
Examples
The following command sets the output current range to 1 mA on channel 1.
CURR:RANG R1mA, (@1)
The following query returns the programmed current range on channel 1.
CURR:RANG? (@1)
Typical Response: R1mA
See Also
[SOURce:]VOLTage:RANGe
Item Type Range of values Default value
<ch> NR1 1 through 3 Required parameter
<range> CPD – R1uA – 1 μA range
– R10uA – 10 μA range
– R100uA – 100 μA range
– R1mA – 1 mA range
– R10mA – 10 mA range
– R120mA – 120 mA range
R1uA