STATus Subsystem 10
Keysight U2722A/U2723A Programmer’s Reference 93
STATus:QUEStionable:CONDition?
Syntax
STATus:QUEStionable:CONDition?
This query returns the value of the Questionable Status Condition register. The
Condition register is a read-only register, which holds the live (unlatched)
questionable status of the instrument. Reading the Questionable Condition
register does not clear it.
Remarks
The following table lists the bit definitions for the Questionable Status registers.
Returned Query Format
<NR1>
The returned value is the binary-weighted sum of all bits set in the register. For
example, if bit 4 is set, this query command will return “+16”.
Example
The following query command reads the Questionable Status Condition register.
STAT:QUES:COND?
Typical Response: +16
For more information on the SCPI Status System, refer to Chapter 2, “Status
system diagram” on page 21.
Bit number Decimal value Definition
0 to 3 Not Used 0 0 is returned.
4 OT 16 The over-temperature protection has tripped.
5 to 15 Not Used 0 0 is returned.