10 STATus Subsystem
82 Keysight U2722A/U2723A Programmer’s Reference
STATus:OPERation:CONDition?
Syntax
STATus:OPERation:CONDition?
This query command returns the value of the Operation Status Condition register.
The Condition register is a read-only register, which holds the live (unlatched)
operational status of the instrument. Reading the Operation Condition register
does not clear it.
Remarks
The following table lists the bit definitions for the Operation Status registers.
NOTE
For more information on the SCPI Status System, refer to Chapter 2, “Status
system diagram” on page 21.
Bit number Decimal value Definition
0 to 1 Not Used 0 0 is returned.
2 DTG1-tran 4 The transient system has received a trigger signal and operation is running in
channel 1.
3 DTG2-tran 8 The transient system has received a trigger signal and operation is running in
channel 2.
4 DTG3-tran 16 The transient system has received a trigger signal and operation is running in
channel 3.
5 WTG1-tran 32 The transient system is waiting for a trigger in channel 1.
6 WTG2-tran 64 The transient system is waiting for a trigger in channel 2.
7 WTG3-tran 128 The transient system is waiting for a trigger in channel 3.
8 to 15 Not Used 0 0 is returned.