9 SOURce Subsystem
78 Keysight U2722A/U2723A Programmer’s Reference
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]
Syntax
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <voltage>,
(@<ch>)
This command sets the voltage trigger level of the specified output channel. Units
are in voltage. The triggered level is a stored value that is transferred to the output
when an output step is triggered.
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? (@<ch>)
Query the stored value of triggered voltage.
Parameters
Returned Query Format
<NR3>
Examples
The following command sets the voltage trigger level to 0.02 V on channel 1.
VOLT:TRIG 0.02, (@1)
The following query returns the voltage trigger level on channel 1.
VOLT:TRIG? (@1)
Typical Response: +2.000000E-02
See Also
INITiate[:IMMediate]:TRANsient
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
[SOURce:]CURRent:RANGe
Item Type Range of values Default value
<ch> NR1 1 through 3 Required parameter
<voltage> NRf The maximum value is dependent on the voltage
range set
Required parameter

Table of Contents