2.4 Filter system
There are atotal of three filters integrated in the Suprasorb
®
CNP P3: An internal filter inside
theSuprasorb
®
CNP P3 device and two external filters in the exudate pouch.
The external charcoal/bacterial filter in the exudate pouch prevents oversuction. Draining stops
when the exudate pouch needs to be changed. The second filter is aventilation filter.
The filter inside the Suprasorb
®
CNP P3 is also acharcoal/bacterial filter and protects the
device against oversuction and contamination.
Should the internal filter become blocked, the Suprasorb
®
CNP P3 must be properly
reprocessed by the L&R service centre or an L&R authorised service partner.
The activated charcoal filter in the container prevents the development of odours.
2.5 Gelling agent
The exudate drained into the exudate pouch thickens within afew minutes due to the
gellingagent contained in the exudate pouch. To avoid leaks– for example, when changing
theexudate pouch– the tube clamp on the exudate pouch tube should be closed off.
The exudate pouch incl. exudate pouch tube and tube clamp is intended
for single use only.
PLEASE NOTE
The internal charcoal/bacterial filter is not intended for reuse and must be
replaced if it comes into contact with the exudate (blockage), once the filter
service life has elapsed or during servicing/repair work performed by L&R
oran L&R authorised service partner.
2.6 Battery
Battery, rechargeable 10.8V, 2.7Ah – lithium ion
Charging time when battery is empty 6-7 hours
Charging time with approx. 50% residual charge 3-3.5 hours
Operating time With battery: 10-20 hours