155
9 Technical data of the Suprasorb
®
CNP P3
therapy unit and EMC information
Aggregate suction power 11 l/min
Negative pressure -20 to -200mmHg*;
conversion factor: 1kPa≈7.5mmHg
Container Suprasorb
®
CNP Container 250ml
Suprasorb
®
CNP Container 1,000ml
Power supply rated voltage
Maximum load current
Power supply line frequency
100-240VAC primary / 15VDC secondary
1A
50/60Hz
Power consumption 15W (charging and operation)
Battery, rechargeable 10.8V, 2.7Ah – lithium ion
Charging time when battery is empty
Charging time with approx. 50%
residualcharge
6-7 hours
3-3.5 hours
Dimensions (H x W x D) 175x155x70mm with Suprasorb CNP Container
250ml
175x260x70mm with Suprasorb CNP Container
1,000ml
Weight (therapy unit) 1.15kg with Suprasorb CNP Container 250ml
1.38kg with Suprasorb CNP Container 1,000ml
Operating time Mains: Continuous operation
Battery: approx. 10-20 hours
Degree of protection in accordance
withEN 60601-1
Type BF
Risk class as per 93/42/EEC,
Annex IX, Rule 11
II
a
Protection class in accordance
withEN60601-1
II
CE marking CE0123
Noise emissions 40dB (A)
Ambient conditions Transport/storage: -10°C to +60°C
Operation: +5°C to +40°C
Recommended charging temperature: +15C to +30C
5 to 80% relative humidity, non-condensing
Atmospheric pressure: 860hPa to 1,060hPa
REF 31780
34780
* Maximum negative pressure fluctuations:
The maximum value refers to sea level.
It may vary depending on altitude and atmospheric pressure. The multiplication factors listed in the table are employed
toset the maximum negative pressure based on the altitude and atmospheric pressure.