Location
0 metres
above sea level
5
00 metres
above sea level
1,000 metres
above sea level
2,000 metres
above sea level
Multiplication factor 1.0 0.94 0.88 0.78
Table 1
Guidelines and manufacturer’s declaration – Electromagnetic emissions
The Suprasorb
®
CNP P3 therapy unit is intended for operation in an ELECTROMAGNETIC
ENVIRONMENT as described below.
The customer or user of the Suprasorb
®
CNP P3 therapy unit must ensure that it is operated
insuch an environment.
Measurement of
emitted interference
Conformity Electromagnetic environment – guidance
HF emissions
in accordance
with CISPR 11
Group 1
The Suprasorb
®
CNP P3 therapy unit only
usesHF energy for its internal workings.
As such, its HF emissions are very low and
itisunlikely that adjacent electronic devices
willbeaffected.
The Suprasorb
®
CNP P3 therapy unit is intended
for use in all facilities including residential buildings
and those directly connected to the mains power
grid which also supplies residential buildings.
HF emissions
in accordance
with CISPR 11
Class B
Harmonic current
emissions in accordance
with IEC 61000-3-2
Class A
Emission of voltage
fluctuations/flickers
inaccordance
withIEC61000-3-3
Conforms