157
Table 2
Guidelines and manufacturer’s declaration – Electromagnetic immunity
The Suprasorb
®
CNP P3 therapy unit is intended for operation in an electromagnetic environment
as described below.
The customer or user of the Suprasorb
®
CNP P3 therapy unit must ensure that it is operated
insuch an environment.
Immunity tests IEC 60601
testlevel
Conformity level Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
inaccordance with
IEC 61000-4-2
± 6kV contact
discharge
± 8kV air
discharge
± 6kV contact
discharge
± 8kV air
discharge
Floors should be made of wood or
concrete or covered with ceramic tiles.
If the floor is covered with synthetic
material, the relative humidity must
bemin. 30%.
Electrical fast
transient/bursts
immunity in
accordance with
IEC 61000-4-4
± 2kV for power
cables
± 1kV for input and
output cables
± 2kV for power
cables
± 1kV for input and
output cables
The quality of the supply voltage should
correspond to that of atypical business
or hospital environment.
Surges in
accordance with
IEC 61000-4-5
± 1kV voltage
phase-phase
± 2kV phase-earth
± 1kV voltage
phase-phase
± 2kV phase-earth
The quality of the supply voltage should
correspond to that of atypical business
or hospital environment.
Voltage dips, short
interruptions and
voltage variations
inaccordance with
IEC 61000-4-11
<5% UT (>95%dip
of the UT)
for0.5cycle
40% UT (60%dip
of the UT)
for5cycles
70% UT (30%dip
of the UT)
for25cycles
<5% UT (>95%dip
of the UT)
for5seconds
<5% UT (>95%dip
of the UT)
for0.5cycle
40% UT (60%dip
of the UT)
for5cycles
70% UT (30%dip
of the UT)
for25cycles
<5% UT (>95%dip
of the UT)
for5seconds
The quality of the supply voltage should
correspond to that of atypical business
or hospital environment.
If the user of the Suprasorb
®
CNP P3
therapy unit requires it to continue
functioning even if the power supply is
interrupted, we recommend connecting
the device to apower supply without
interruptions or running it off the
battery.
Magnetic fields
at the supply
frequency
(50/60Hz) in
accordance with
IEC 61000-4-8
3A/m 3A/m
Magnetic fields at the supply frequency
should be equivalent to the typical
values encountered in business and
hospital environments.
NOTE: UT is the mains alternating voltage prior to application of the test level.