3.3 Insertion of the Suprasorb
®
CNP Exudate Pouch
and connection of the container system
CAUTION
Exudate pouch
Ensure that the exudate pouch for the Suprasorb
®
CNP P3 is replaced
regularly. When the exudate pouch is full, the integrated overflow prevention
system is triggered. This discontinues the therapy.
Switch the device to Pause mode when replacing the exudate pouch.
PLEASE NOTE
Collapsing exudate pouch
A leak in the container or the cover of the exudate pouch may result
inairentering the container. This, in turn, may cause the exudate pouch
tocollapse.
When assembling the container system, ensure that the side plate of
theexudate pouch is securely connected to the container.
Ensure that all connections are secure and established correctly.
Ensure that neither the container nor the exudate pouch is damaged.
The separate instructions for use for the exudate pouch must also be
followed!
The container system of the Suprasorb
®
CNP P3 comprises the Suprasorb
®
CNP Container,
the Suprasorb
®
CNP Exudate Pouch with the connected exudate pouch tube and atube clamp.
There are two possible combinations here:
A) Suprasorb
®
CNP
ExudatePouch 250ml
andSuprasorb
®
CNP
Container 250ml
B) Suprasorb
®
CNP
ExudatePouch 1,000ml
andSuprasorb
®
CNP
Container 1,000ml