153
8 Transport, storage and disposal
Please read “Important information on your safety”, Section 1.5.
8.1 Decontamination before shipping
CAUTION
Avoiding infections
Infectious and pathogenic microbes in the exudate can pose health risks.
Wear appropriate disposable gloves when replacing the container.
Replace the exudate pouch including the exudate pouch tube in
accordance with the respectively applicable hygiene regulations, at the latest
every week, before achange of patient or when full.
Reprocessing is required whenever there is achange of patient.
Reprocessing means that all the single-use products are disposed of
and the Suprasorb
®
CNP P3 therapy unit wipe-down disinfected. The
Suprasorb
®
CNP P3 therapy unit must only be opened by L&R or an
authorised L&R service partner.
The exudate and all components contaminated with it must be disposed of
in accordance with the facility’s policies.
CAUTION
Use of the carry bag
The Suprasorb
®
CNP Carry Bag for mobile use supplied with the
Suprasorb
®
CNP P3 is intended for single use per patient. It should not be
used for other patients but be disposed of once used.
CAUTION
Shipping instructions
The therapy unit must be sent in the supplied case, which must also be
protected by an additional box.
Used exudate pouches must not be sent with the device but should be
disposed of properly.
Before the Suprasorb
®
CNP P3 therapy unit is passed on to the next user, it must be
professionally reprocessed by L&R or an authorised specialist.
The reprocessing must be performed in accordance with the Medical Devices Operator Ordinance
(MPBetreibV), the Medical Devices Act (MPG) and the manufacturer’s specifications. L&R offers
its customers and partners rapid and professional reprocessing as well as the performance of
necessary tests and checks.
The Suprasorb
®
CNP P3 therapy unit must be cleaned and disinfected before dispatch
totheL&R Service Centre. Please refer to Section 7.
8.2 Storage
Store the Suprasorb
®
CNP P3 therapy unit in accordance with the specifications
inTechnical Data (Section 9).
Charge the Suprasorb
®
CNP P3 therapy units battery before storing the device.
Thisensures that it can be used at any time.
If the Suprasorb
®
CNP P3 therapy unit is not used for an extended period of time
(severalmonths), the battery should be fully recharged!