139
“Continuous” or “Intermittent”
In “Continuous” mode you set aconstant
negative pressure for the therapy.
In“Intermittent” mode, you select anegative
pressure which alternates between twovalues.
In this case, you need to specify amaximum
and minimum negative pressure and set the
duration of the respective setting in minutes
in the time bar (under the pressure setting
barrespectively).
Set the desired strength
ofthe negative pressure in
theline with the pressure
symbol
. In addition,
setthe time for which the
selected pressure should be
retained in the line with the
clock symbol
using the
keys.
4.7 Setting the sensitivity
WARNING
In case of low/minimal
sensitivity
The sensitivity setting is
important for ensuring that the
message system is tailored to
the patient’s personal wound
situation. The sensitivity of the
Suprasorb
®
CNP P3 canbe
setto “High”, “Medium”,
“Low”or “Minimal”.
• However, the sensitivity must
only be set by aqualified and
authorised specialist.
• If the sensitivity is lowered
considerably, the risk of
ablockage or leak in the
wound and tube system
goingundetected and
noalarm being triggered
cannot be excluded.
• As such, if the sensitively
is lowered considerably,
particularly careful
monitoring of the therapy
must be ensured.
CAUTION
Concerns
qualified
healthcare
professionals
Setting the therapy
parameters
• Before using the
Suprasorb
®
CNP P3 on
anew patient, first check
thatthe therapy parameters
set are suitable for the patient.