LR Suprasorb CNP P3 - Setting the sensitivity

LR Suprasorb CNP P3
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
139
“Continuous” or “Intermittent”
In “Continuous” mode you set aconstant
negative pressure for the therapy.
In“Intermittent” mode, you select anegative
pressure which alternates between twovalues.
In this case, you need to specify amaximum
and minimum negative pressure and set the
duration of the respective setting in minutes
in the time bar (under the pressure setting
barrespectively).
Set the desired strength
ofthe negative pressure in
theline with the pressure
symbol
. In addition,
setthe time for which the
selected pressure should be
retained in the line with the
clock symbol
using the
keys.
4.7 Setting the sensitivity
WARNING
In case of low/minimal
sensitivity
The sensitivity setting is
important for ensuring that the
message system is tailored to
the patient’s personal wound
situation. The sensitivity of the
Suprasorb
®
CNP P3 canbe
setto “High, “Medium”,
“Low”or “Minimal.
However, the sensitivity must
only be set by aqualified and
authorised specialist.
If the sensitivity is lowered
considerably, the risk of
ablockage or leak in the
wound and tube system
goingundetected and
noalarm being triggered
cannot be excluded.
As such, if the sensitively
is lowered considerably,
particularly careful
monitoring of the therapy
must be ensured.
CAUTION
Concerns
qualified
healthcare
professionals
Setting the therapy
parameters
Before using the
Suprasorb
®
CNP P3 on
anew patient, first check
thatthe therapy parameters
set are suitable for the patient.