4.3 Start screen
The start screen appears
forafew seconds when
theunit is switched on.
4.4 Preset therapy mode
The preset therapy mode
appears after the start
screen; the therapy
isstopped.
4.5 Changing settings
The screen of the Suprasorb
®
CNP P3 is
atouchscreen. The following changes can be
made here by simply touching the symbols in
the two toolbars (see red marking in image of
screens at end of Section 4.5):
Start of therapy
The therapy mode must
onlybeset by qualified
healthcare professionals!
General settings
Locking function
4.6 Setting the therapy mode
Selecting takes you
tothe following screen:
You can cancel the
action by pressing
orconfirm your selection
by holding
down for
3seconds to proceed to
thetherapy settings menu.
Choose one of the two
therapy modes by selecting
the corresponding view –
Continuous or Intermittent–
using
. The values
are set in intervals of 5
forthe pressure and
30seconds for the time.