Using the power socket on the Suprasorb
®
CNP P3, connect the unit to the mains power
supply for charging or operation via the power supply.
PLEASE NOTE
Incorrect voltage power supply
Improper operation causes overvoltages to develop in the device which
canthen pass to the user.
• Ensure that the power supply is set to avoltage of 100-240V alternating
current prior to use.
• Only operate the device with the power supply TR15RAM150-E-160E24
input 15V DC 1A.
If required, there is an adapter supplied in the case with which the device
canbe connected up to the mains power supply in your country.
Use the supplied power supply only.
Start by connecting the power supply to the power socket on the Suprasorb
®
CNP P3.
Then connect the power supply to the mains power supply.
Connection of
Suprasorb
®
CNP P3
therapy unit to patients
and accessories
Power cable
Power supply
of Suprasorb CNP P3
Socket
Exudate pouch tube Wound drainage system
Container system
(container &
exudate pouch)
Device
for generation
of negative pressure
Patient