Error message Status Possible cause Remedy
A signal sounds.
Therapy is
interrupted.
Container is not
connected.
Connections open.
The container and exudate pouch may
not be connected to the device properly.
The cause may also be acollapsed
exudate pouch.
Or aconnection between the tubes
maybe open.
Please check the points above and
re-establish all connections if necessary.
A signal sounds.
Therapy continues
to run in the
background.
Wound dressing
or connections
nottightly sealed.
The tubes might not be connected
properly or the dressing itself is not
sealed tight.
Please check both possibilities,
reconnect the tubes using
thesingle-piece connector or
improvetheseal onthe dressing.
A signal sounds.
Therapy continues
to run in the
background.
Exudate flow
interrupted, tube
blocked, internal
filter blocked.
A tube clamp may have accidentally
been closed or there may be ablockage
in the tubes.
Please open all the tube clamps or try
toremove the blockage in the tubes.
If necessary, change the full exudate
pouch.
However, if the internal charcoal/bacterial
filter is blocked, the device needs to
be reprocessed properly by L&R or
anauthorised L&R service partner!
A signal sounds.
Therapy continues
to run in the
background.
If the message “Exudate pouch is full”
appears, it is high time to insert anew
exudate pouch. Replace the pouch in
accordance with the instructions for use
(see Section 3.4).