50
Mahr GmbH • Millimar C1202
5.1.5 Number of masters
– Select the Number of masters
submenu using the ▼ or ▲ key.
– Select using 4 key or exit submenu
using the Ó key.
– Select the number of masters using
the ▼ key.
1 = one-point master measurement
2 = two-point master measurement
see Section 3.2.2. Master
– Use the 4 key to conrm the
selection or Ó to exit the submenu.
– Use the
key to switch to
measuring mode.
5.1.6 Master reset
– Select the Master reset submenu
using the ▼ or ▲ key, or exit the
submenu using the Ó key
or switch to measuring mode using
the
key.
– Use the 4 key to delete the master
values set in measuring mode. C
1202 switches to measuring mode.
The values displayed now refer
directly to the probe signal.
Master value is set Master value is
deleted