Mahr GmbH • Millimar C1202
5.2.6 Start delay
Select the Start delay submenu
using the or key.
Select using 4 key or exit submenu
using the Ó key.
Use the
key to select the digit and
the or key to enter the relevant
value.
Use the 4 key to conrm the value or
Ó to exit the submenu.
If you want a certain amount of
time to pass between the moment
(Start) is pressed and the
measurement starts, you can enter
a time period by which the start
of the measurement is delayed.
A start delay is particularly useful
when a dynamic measurement (e.g.
on a shaft) is performed and it is
supposed to start automatically
via a switch contact on the
measuring device. When the shaft
is approached, vibrations occur in
the measuring device that affect the
measuring result. During the delay
time, the vibrations subside and no
longer affect the measuring result.