Mahr GmbH • Millimar C1202
The instrument is most accurate after approx. 30 minutes of operation. To perform a
reliable customer calibration using a probe, the probes electrical zero point should be
positioned at the calibration zero point using the probes raw value display. (see point
6.1.1.).
This is because the inductive probe measures most accurately at the electrical zero
point. The probes electrical zero point should be set on a 1.000 mm gage block for
a customer calibration, like the one in this example, where a 2.000 mm gage block is
used for the MAX value and the measuring plate represents the MIN value.
6.1.2.1.2 Performing customer calibration
Select the Calibration submenu
using the or key.
Select using 4
Select the Calibration C1 submenu
using the or key.
Select using 4
Select Perform customer calibration
using the or key.
Select using 4 key or exit submenu
using the Ó key.
– Use
to select the value for
Calibration min.
Use the 4 key to select the
Calibration min submenu.