EasyManua.ls Logo

marposs TESTAR E4N - Rogramming Error Messages and Solutions

marposs TESTAR E4N
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
82
10
********
________
________
????????
ERR 6
ERR 6






ERR 20
PROGRAMMING ERROR MESSAGES AND SOLUTIONS
Programming
Column waiting for
programming to be
completed
Please wait
Hardware
Sensor not connected
Measurement out of
range
Connect the sensor.
Check that the sensor
is touching the part.
Program another measu-
rement range.
Programming
Data value not program-
med
Program the value
Programming
Column not program-
med
Program the column
Programming
The distance between
the MIN and MAX ma-
sters is too small
Check that the correct
masters are being used
or program the corrects
tolerance limits
MEASURE:
Programming
N.MEAS=2 M.TYPE=
not programmed
or

not programmed
Program
M.TYPE=STATIC
ADVANC.:
Programming
Measurement system
not programmed
Program
M.TYPE=STATIC
or
MAX, MIN, (M+m)/2,
(M-m)/2 or (M-m)
MEASURE:
Programming
Programming error in
LANGUAG= parameter
Program
LANGUAG=ENGLISH or

DEUTSCH, ESPANOL
ADVANC.:
Programming
System of measurement
(metric or imperial) and
unit of measurement
(mm, micron, inch, ecc.)
are incompatible
Program
M.SYST= and M.UNIT=
so that they are com-
patible
MEASURE:
/
ADVANC.:
Programming
N.MEAS= not program-
med

MEASURE:
Programming
IDENT.= incorrectly
programmed
Program
IDENT.=
ADVANC.:
Programming
Hardware
Amplifier module incom-
patible with program-
ming (no. sensors or
inputs are incompatible)
Modify the program so
that the number of inputs
or sensors is correct, or
install a different amplifier
module
MEASURE:
ACTIONMENU
ALPHANUMERIC
DISPLAY
7-SEGMENT DI-
SPLAY
ERROR TYPE DESCRIPTION

Table of Contents

Related product manuals