3.8.11
3.8.10
MASTER 2= (Deviation of master gauge 2)
zero the measurement and this function (Master 2) can be used to correct sensitivity.
Step 1: Press the ENTER key to set the variable.
Step 2:
digit is flashing). Use the UP and DOWN arrow keys to increase or decrease the value
of the flashing digit.
Note: Not to program (Reset) the variable, repeatedly press the UP or DOWN arrow key when
the first digit on the left of the display is flashing.
Step 3: Press the ENTER key to confirm the numeric data.
MASTER 2=
Step 1
-0011.0
Step 3
Step 2
R____._
-0015.0
+9999.9
MASTER 1= (Deviation of master gauge 1)
Step 1: Press the ENTER key to set the variable.
Step 2:
DOWN arrow keys to increase or decrease the value of the flashing digit.
Step 3: Press the ENTER key to confirm the numeric data and to pass to programming of the next variable.
MASTER 1=
Step 1
Step 2
Step 3
+0015.0
-0015.0
+0015.0
HOW TO PROGRAM THE E4N COLUMN