EasyManuals Logo

Mecal CDA10 User Manual

Mecal CDA10
119 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
CRIMP DATA ANALYZER
CDA10
REV.01
58
Connect the yellow cable to the pos.6 input on the terminal board;
Connect the green cable to the pos.7 input on the terminal board;
Close the door on the electrical box;
Connect the press to the electrical mains;
Move the switch to the ON (I) position;
Configure the load cell [see the relative paragraph].
5.2.4 REPLACING THE FORCE CELL TT1000 WITH CDA10
The operations to be carried out to replace load cell TT1000 with load cell CDA10 are:
Check that the switch, located on the press or in the electrical cabinet/panel, is in the OFF (O)
position;
Disconnect the press from the electrical mains;
Disassemble cell TT1000 to be replaced, disconnecting it from all connectors;
The load cell sensor does not need to be disassembled (if working);
The encoder does not need to be disassembled (if working);
Install the new sensing unit on its support [see relative paragraph];
Connect the load cell sensor to the relative input;
Connect the encoder to the load cell sensing unit
using the supplied adapter (Z);
Connect the cables from the sensing unit to the
press, following the paragraph on connection
referring to the corresponding press model;
The new load cell must be configured [see relative
paragraph].

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mecal CDA10 and is the answer not in the manual?

Mecal CDA10 Specifications

General IconGeneral
BrandMecal
ModelCDA10
CategoryMeasuring Instruments
LanguageEnglish