105
Endoscope Errors
Code Description
S-102 Start not allowed: vent hookup block
S-103 Endoscope leak or connected incorrectly
S-104 Minor endoscope leak, cycle continuing safely
S-105 Leak test vented but pressure remains high
S-106 The block is not pressurized or is not connected
S-112 Start not allowed: vent hookup block
S-113 No ow: air channel
S-114 Blocked air channel
S-115 No pressure: air channel
S-116 Air channel disconnected
S-117 Air channel disconnected
S-122 Start not allowed: vent hookup block
S-123 No ow: biopsy channel
S-124 Blocked biopsy channel
S-125 No pressure: biopsy channel
S-126 Biopsy channel disconnected
S-127 Biopsy/Suction channel disconnected
S-132 Start not allowed: vent hookup block
S-133 No ow: jet channel
S-134 Blocked jet channel
S-135 No pressure: jet channel
S-136 Jet channel disconnected
S-137 Jet channel not connected
S-142 Start not allowed: Vent hookup block
S-143 No ow: suction channel
S-144 Blocked suction channel
S-145 No pressure: suction channel
S-146 Suction channel disconnected
S-147 Suction/biopsy channel disconnected
S-152 Start not allowed: Vent hookup block
S-153 No ow: water channel
S-154 Blocked water channel