119
ADVANTAGE PLUS System manager.
Channel disconnected
S-116, S-126, S-136,
S-146, S-156, S-166,
S-176
This control checks interconnected endoscope channels and the message is given per channel, i.e.:
• Air • Suction • Elevator
• Biopsy • Jet
• Extra • Water
Cause: There is a connector loose from the endoscope.
Action: If you see a loose connection, press the Stop
button. Close the program, connect the loose
connection, and restart the cycle.
Cause: It is not clear if all channels are connected.
Action: Try to start the process again by pressing the Start
button. This may be tried once. If the
message reoccurs, you must press the Stop
button and call Maintenance.
Channel not disconnected
S-117, S-127, S-137,
S-147, S-157, S-167,
S-177
This control checks small endoscope channels and the message is given per channel, i.e.:
• Air • Suction • Elevator
• Biopsy • Jet
• Extra • Water
Cause: There is a connector loose from the endoscope.
Action: If you see a loose connection, press the Stop
button. Close the program, connect the loose
connection, and restart the cycle.
Cause: It is not clear if all channels are connected.
Action: Try to start the process again by pressing the Start
button. This may be tried once. If the
message reoccurs, you must press the Stop
button and call Maintenance.
Reprocessing Cycle Errors
Internal machine uid leak, check drip tray S-201
If the message report is visible, then the program can not be stopped. After drying out, it can take 5–10
minutes before the sensor message disappears. If the sensor is released, only the program can be stopped.
Cause: There is liquid in the front of the machine as result of a leak.
Action: If possible, close the water supply tap and consult the maintenance department
Cause: If the tray is dry, a sensor is out of alignment.
Action: Remove the two containers and dry the green sensor with a cloth. If the message remains, report
the error to the maintenance department to have the sensor aligned, or to have the program shut