down for technical service.
Wrong connection block S-203
The machine disinfection automatically started and checked whether the hot water block is tted.
Cause: No connection block is present:
Action: Place the machine disinfection connection block and then press the Start
button.
Cause: The endoscope connection block is present.
Action: Exchange this block for a machine disinfection connection block and press the Start
button.
Cause: The machine disinfection connection block is present.
Action: Try pressing the Start
button again. If the message recurs, then consult the maintenance
department.
Lid open S-211
Cause: The lid is closed: During the program, the lid sensor has failed and now indicates that the cover
is open.
Action: The program can be stopped only via the expert operation. Report the error to the maintenance
department to adjust the sensor, if necessary.
Cause: During the program the lid was opened.
Action: Close the lid and press the Start
button to continue the program.
Start not allowed: Check detergent pressure sensor S-222
Cause: The detergent sensor is out of alignment or faulty.
Action: Report the fault to the maintenance department.
No detergent ow S-223
Cause: The detergent reservoir is empty.
Action: Check if the detergent reservoir is empty. If so, ll the reservoir. If there is sufcient detergent
in the reservoir, press the Start
button. If the error repeats, see the alternate cause below.
Cause: The detergent sensor or pump is defective.
Action: Report the fault to instrument service.
Start not allowed : Detergent level low.
Start not allowed : Alcohol level low.
S-224,S-244
Cause: The reservoir is empty.
Action: Check if the level of uid in the reservoir. If it is empty, ll or replace the reservoir. If there is
sufcient uid in the reservoir, press the Start
button. If the error repeats, see the alternate
cause below.
Cause: The sensor in the reservoir is defective.