5. When the designated start time is reached, a system check is carried out. If a yellow error report appears
at the bottom of the screen, attempt to correct the error by consulting Chapter 5: Troubleshooting.
6. The reprocessing cycles is initiated at the selected start time.
Cycle Interruptions
Introduction
At startup and during the disinfection program cycle, the MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope Reprocessor monitors process parameters for the endoscope, machine, and software
to ensure accurate disinfection of the endoscope. Reports, program errors, and system errors appear on the
screen to communicate the cycle progress to you.
Operator messages and program errors appear at the bottom of the screen in the Reports eld. A code
number and short description of the report or error is shown. The elapsed time at which the message or
error is generated is also provided.
System errors appear at the top of the screen in white letters on a red background.
Detailed error reports and troubleshooting steps are provided in Chapter 5.
WARNING If system power is lost during a reprocessing cycle, the system will not complete the
reprocessing cycle once the power is restored. The endoscope is not disinfected and should
not be used on a patient. You must start the reprocessing cycle over so that the endoscope is
properly disinfected.