86
Procedure Summary
1. Open the cover.
2. Connect the endoscope.
3. Close the cover.
4. Enter your login name and password, the endoscope ID, the patient number, the physician ID, and the
assistant ID.
5. Start the program.
6. After disinfection, open the basin lid.
7. Enter your login name and password.
8. Check the sprayer and the connections.
9. Disconnect the connector block.
10. Remove the endoscope from the basin.
11. Print a report.
Downtime
Brief Downtime
• If the MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope Reprocessor will be unused for
a brief period time—less than seven days—it should remain powered on.
• To prevent moisture in the leak test system, t disinfection blocks in the MEDIVATORS ADVANTAGE
PLUS System when it is not is use.
Prolonged Downtime
• If the MEDIVATORS ADVANTAGE PLUS System will be unused for more than seven days, it should
be powered off.
• The tubing, blocks, and lters in the machine contain a lot of water where bacterial growth may occur.
After a long downtime, replace the bacterial lter and disinfect the MEDIVATORS ADVANTAGE
PLUS System before running any endoscope disinfection procedures.
• To shut down the MEDIVATORS ADVANTAGE System, close all running programs rst before
pressing the main PC power switch. Failure to do this may result in software errors.
• See storage section for further information.