External water supply valve failed S-289
Cause: The water supply valve on the lter unit is leaking, so machine disinfection can not be performed.
Action: Report the error to the maintenance department.
Basin temperature sensor defective S-292
This error occurs independently of the software and is an extreme situation. Pressing the Stop
button
has no effect. Interrupting the program is only recommended after a technician has judged the situation. The
program can be stopped via the expert operation. After resolving the fault, the program can be continued.
Cause: The temperature sensor is defective if the temperature message on the screen is higher than
100°C.
Action: Report the error to the maintenance department.
Basin temperature too high S-293
Press the Start
button to start the process step again. If the message recurs, press the Stop
button.
Cause: The drainage block is too warm after a disinfection program.
Action: Wait ve minutes, or pour cold water in the rinse basin to see if the temperature indicated on
the screen drops.
Cause: The setting or mixing valve is out of order, causing the supplied water to be too hot.
Action: Press the Stop
button and check or adjust the setting of the mixing valve. Wait for a while
until the water in the machine has cooled down. Then start the program again.
Basin temperature too low S-294
Press the Start
button to start the process step again. If the message recurs, press the Stop
button.
Cause: The adjustment or mixing valve is out of order, causing the water supply to be too cold.
Action: Contact the maintenance department. Press the Stop
button and check or adjust the mixing
valve. Start the program and check if the temperature indicated on the screen is correct for the
program.
Cause: The warm water supply is stagnated.
Action: Press the Stop
button and consult the maintenance department.
Basin temperature too high S-297
This error occurs independently of the software and is an extreme situation. Pressing the Stop
button has no effect. Interrupting the program is only recommended after a technician has judged the
situation. The program can be stopped via the expert operation. After resolving the fault, the program can
be continued.
Cause: The water temperature in the rinse basin is too high:
Instrument disinfection: >50°C.
Machine disinfection: >90°C.
Action: Stop the program and report the fault to the maintenance service.