Measured contact time is too short S-511
The prescribed contact time is a sum of the time for different steps
(SV + SS + SC + RS = option).
Cause: There was a change in the drain and ll time for the basin.
Action: Report the error to Maintenance so they can recalibrate.
Cause: The program parameters have been changed.
Action: Report the fault to the system administration.
MRC validation failed S-660
Cause: At the end of the cycle, the operator determined the disinfectant sample failed the MRC
validation test and pressed the stop button as directed.
Action: Replace the part A and part B containers and rerun the cycle.
“Machine Control” Locked by the other side S-671
Cause: The machine disinfection program is active. It is not possible to use this side of the machine.
Action: Wait until the program has run completely.
Wrong system side S-672
Cause: The machine disinfection program is started from the right side rather than the left side.
Action: Cancel the selection from the right side and restart it on the left side.
Air cylinder does not charge S-831
Cause: The cylinder is used to check connectivity from the endoscope to the MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope Reprocessor. The cylinder is not lled so the
check can not be performed.
Action: This is the last step of the disinfection cycle. When all channels are still connected, the endoscope
is properly disinfected. Visually check the connection and report the error to maintenance.
Air cylinder not discharged (cylinder pressurized) S-832
Cause: The cylinder is used to check connectivity from the endoscope to the MEDIVATORS
ADVANTAGE PLUS system. The cylinder can not be vented after the check is performed.
Action: Report the error to maintenance.