Daily Maintenance
Details and instructions for the daily service requirements are included in Chapter 3: Operation.
The following daily maintenance procedures must be completed at the start of each day:
• Conrm that the main air pressure regulator is set between 2.8–3.0 bar (42.0 – 45.0 psi).
• Conrm that the leak test air pressure regulator is set at 0.24–0.25 bar (3.6 – 3.75 psi).
• Verify a sufcient ow of water into the basin or replace the system lters.
• Check that all O-rings on the connection columns are present and undamaged.
• Replace the screens in the rinse basin if they are soiled.
At the end of each day, clean the MEDIVATORS
®
ADVANTAGE PLUS
®
system:
• For the top and outside covers, use a non-scouring cleansing agent.
• On the endoscope basin and inside covers, use a non-scouring, lime-dissolving cleansing agent.
• Use soap and water to clean the cover gaskets and prevent them from sticking.
Weekly Maintenance
Inspection
Inspect the following on a weekly basis. Adjustments or changes should be made by a qualied service
technician.
• Hookup Connector Blocks: Check for cracks in the hookup hoses, especially at the connections
to the connector block. If cracks have formed, the hose should be replaced.
• Glass Covers: The basin lids should open and close easily.
• Door: The front door should open and close easily.
• MEDIVATORS ADVANTAGE PLUS System Tubing, Connections, Fittings, Hydraulics
and Dosing System: Check for wear and leakage and replace as needed.