Warning Summary
WARNING Portable and mobile communication devices can affect electrically-operated medical
equipment
WARNING Connecting a chemical container with the incorrect uid may cause the endoscope to
be improperly disinfected. Do not use the endoscope on a patient if this occurs.
WARNING Wear personal protective equipment (PPE) (clothing, mask, gloves & eye wear).
WARNING All endoscope hookups must be periodically inspected for wear. Discard and replace
damaged hookups otherwise it could result in an endoscope that is not disinfected and therefore
should not be used on a patient.
WARNING Verify all connections to the endoscope are secure and tubing is not linked. Failure
to do so could result in an endoscope which is not disinfected and therefore should not be used
on a patient.
WARNING Verify the connection tubing does not interfere with the spray head and conrm that
the spray function is working during the cycle. Failure to do this could result in an endoscope
that is not disinfected and therefore should not be used on a patient.
WARNING Before removing the endoscope, verify all connections to the endoscope are secure.
If an adapter is loose or disconnected, the disinfection was not complete and must be repeated.
Failure to do so could result in an endoscope which is not disinfected and therefore should not
be used on a patient.
WARNING Report all system errors to your supervisor.
WARNING Wear clean gloves to prevent soiling of the disinfected endoscope.
If a run error was generated, a channel is blocked, or the endoscope is leaking, the endoscope
must not be released for patient use. If there is any doubt as to whether an endoscope is
correctly disinfected, it must be disinfected again before use. Contact your supervisor for further
instructions.
Verify all endoscope connections are secure. Failure to do so could result in an endoscope which
is not disinfected and therefore should not be used on a patient.
WARNING Avoid possible chemical burns. Wear personal protective equipment (clothing, mask,
gloves & eye wear).