79
Complete the Disinfection Process
When the disinfection process is complete, the process LED indicator light blinks green, the message
“Completed” displays on the LCD screen, and the alarm sounds.
Table 2 Post-Disinfection Program Status
Status Description
ACTIVE The program is progressing correctly.
CORRECT The endoscope was disinfected without any faults or program interruptions.
CORRECT!!! The endoscope was disinfected with errors reported.
STOP (fault) The program is stopped due to an error.
STOP (ASC) The safety control program stopped the program.
STOP (OPERATOR) The operator stopped the program without a fault report
STOP (LIO) The program stopped automatically after an error report or the maximum
error time (50 minutes) was reached.
SHUTDOWN (fault) After an error, the endoscope is rinsed so that it can be safely removed.
SHUTDOWN (OPERATOR) The operator stopped the program. Run down program.
SHUTDOWN (ASC) The safety control program stopped the disinfection. Completion program.
ASC (REJECT) The safety control program rejected the disinfection. Open the lid to close
the program.
LIO (REJECT) Disinfection was rejected due to an automatic program stop.