80
Table 3 Post-Disinfection Instrument Status
Status Description
DISINFECTED The endoscope was run through a complete disinfection program.
NOT APPLICABLE The program is not an endoscope disinfection program. Warning: The
endoscope can be used only if the connectors are correctly connected
after the program.
NOT SUITABLE FOR USE The endoscope is insufciently disinfected.
WARNING Wear clean gloves to prevent soiling of the disinfected endoscope.
If a run error was generated, a channel is blocked, or the endoscope is leaking, the endoscope
must not be released for patient use. If there is any doubt as to whether an endoscope is
correctly disinfected, it must be disinfected again before use. Contact your supervisor for further
instructions.
Verify all connections to the endoscope are secure. Failure to do so could result in an endoscope
which is not disinfected and therefore should not be used on a patient.
1. Open the basin lid by pressing the Open/Close
button or by using the foot switch. The run status
report disappears and the Operator selection window opens.
2. Enter your login name and password.
3. Check that all adapters are still connected to the channel connections. If an adapter is loose or
disconnected, the disinfection was not complete and must be repeated.
WARNING Before removing the endoscope, verify all connections to the endoscope are secure.
If an adapter is loose or disconnected, the disinfection was not complete and must be repeated.
Failure to do so could result in an endoscope which is not disinfected and therefore should not
be used on a patient.