13
Reprocessor Specications
General Specications
The basic reprocessor specications required for installation and operation of the MEDIVATORS
®
ADVANTAGE PLUS
®
Automated Endoscope Reprocessor are summarized in Table 1 .
Table 1 Reprocessor Specications
Equipment Type Laboratory Equipment
Electrical rating 100-240 VAC, 50/60 Hz
Power rating 1000 Watts
Air pressure range 87 – 145 psi (6 - 10 bar)
Water pressure range 35 - 87 psi (2 - 6 bar)
Water ow range 2 – 8 Liters/minute
Water temperature operating range 30ºC – 40ºC, set mixing valve to 35ºC
Humidity range for operation 20–80%, non-condensing
Temperature range for operation 50°F – 80.6°F (10°C - 27°C)
Environment Indoor use only
Installation/Over voltage category II
Pollution degree 2
Connection to power supply Cord connection
Dimensions
45”W X 31” D X 62” H (inches)
114 W X 79 D X 157 H (cm)
Weight 400 lbs. (182 kg)

Table of Contents

Other manuals for Medivators ADVANTAGE PLUS

Related product manuals